| # SCD41 CO2 Sensor Driver <!-- omit in toc -->
 | |
| 
 | |
| ## Getting Started
 | |
| 
 | |
| Construct a new PimoroniI2C instance for your specific board. Breakout Garden uses pins 4 & 5 and Pico Explorer uses pins 20 & 21.
 | |
| 
 | |
| Since SCD41 has a fixed I2C address and the Sensirion SCD4x library is used under the hood, it's wrapped up as a module for Python.
 | |
| 
 | |
| Import the `breakout_scd41` and call `init` to set up I2C:
 | |
| 
 | |
| ```python
 | |
| import time
 | |
| 
 | |
| import pimoroni_i2c
 | |
| import breakout_scd41
 | |
| 
 | |
| i2c = pimoroni_i2c.PimoroniI2C(4, 5)
 | |
| 
 | |
| breakout_scd41.init(i2c)
 | |
| ```
 | |
| 
 | |
| ## Taking Measurements
 | |
| 
 | |
| Before taking a measurement you must start periodic measurement by calling `start()`.
 | |
| 
 | |
| Poll on `ready()` and use `measure()` to read the result when it's `True`:
 | |
| 
 | |
| ```python
 | |
| breakout_scd41.start()
 | |
| 
 | |
| while True:
 | |
|     if breakout_scd41.ready():
 | |
|         co2, temperature, humidity = breakout_scd41.measure()
 | |
|         print(co2, temperature, humidity)
 | |
|         time.sleep(1.0)
 | |
| ```
 | |
| 
 | |
| The `measure()` method will return a Tuple containing the CO₂ reading, temperature in degrees C and humidity.
 | |
| 
 | |
| ## Changing Calibration
 | |
| 
 | |
| By default the SCD41 will perform automatic self calibration, which could lead to drift in readings over time.
 | |
| 
 | |
| You can stop this with `breakout_scd41.set_automatic_self_calibration(False)`.
 | |
| 
 | |
| You can then use force recalibration with a known good CO₂ PPM baseline to calibrate your sensor:
 | |
| 
 | |
| ```python
 | |
| correction_amount = breakout_scd41.scd41_perform_forced_recalibration(target_co2_concentration);
 | |
| ```
 | |
| 
 | |
| `correction_amount` is the resulting correction in CO₂ PPM.
 | |
| 
 | |
| To successfully conduct an accurate forced recalibration, the following steps must be carried out:
 | |
| 
 | |
| 1. Operate the SCD4x in a periodic measurement mode for > 3 minutes in an environment with homogenous and constant CO₂ concentration. (read: don't breathe)
 | |
| 2. Stop periodic measurement.
 | |
| 3. Wait 500 ms.
 | |
| 4. Issue the perform_forced_recalibration command.
 |